1、Measuring range:0~1000 mg/L(when the BOD5 value exceeds the measuring range, it needs to be diluted)
2、Simultaneous determination of the number of samples: 6 samples or 8 samples each time
3、Accuracy: in line with the national standard "GB7488-87"
(The BOD5 value of the standard solution of glucose and glutamic acid is in the range of 180 mg/L to 230 mg/L)
4、Display: The LCD display cyclically displays the BOD5 value and biochemical reaction curve of each sample.
5、Culture temperature:20℃±1℃
6、Power supply:AC 220V±22V 50Hz±0.5Hz
7、Power consumption:Instrument≤50W,Incubator≤300W
8 、The instrument can save the BOD5 values and biochemical reaction curves of the eight samples analyzed last time